Assessing falls in the elderly population using G-STRIDE foot-mounted inertial sensor
Marta Neira Álvarez1, Antonio R Jiménez Ruiz2, Guillermo García-Villamil Neira2
1Department of Geriatrics, Foundation for Research and Biomedical Innovation of the Infanta Sofía Hospital (HUIS), Madrid, 28702, Spain.
Scientific Reports
|June 6, 2023
Summary
The G-STRIDE system, a wearable sensor, accurately measures gait parameters to identify fall risk in older adults. This technology improves upon existing fall risk assessments, aiding clinical practice.
Area of Science:
- Gerontology
- Biomedical Engineering
- Clinical Biomechanics
Background:
- Falls are a major health concern for the elderly, leading to significant morbidity and mortality.
- Current fall risk assessments often rely on subjective clinical scales and may lack precision in gait analysis.
- There is a need for objective, user-friendly tools to assess gait and balance in daily clinical practice.
Purpose of the Study:
- To clinically validate the G-STRIDE system, a wearable 6-axis inertial measurement unit (IMU), for assessing gait parameters related to fall risk in older adults.
- To evaluate the correlation between G-STRIDE-derived walking metrics and established clinical markers of fall risk.
- To determine if G-STRIDE can effectively discriminate between elderly individuals with and without a history of falls.
Main Methods:
- A cross-sectional case-control study involving 163 participants (falls and non-falls groups).
- Participants underwent clinical assessments and a 15-minute walking test wearing the G-STRIDE system.
- Gait parameters were calculated from IMU data, and correlation analysis was performed with clinical variables.
Main Results:
- The G-STRIDE system demonstrated good to excellent estimation accuracy for walking speed (ICC = 0.885).
- Walking parameters derived from G-STRIDE statistically discriminated between the falls and non-falls groups.
- Gait speed showed good to excellent correlation with several clinical fall risk indicators.
Conclusions:
- The G-STRIDE system provides a low-cost, flexible solution for objective gait analysis in real-world settings.
- G-STRIDE-derived walking metrics can effectively identify individuals at higher risk of falls.
- Preliminary fall-risk assessment using G-STRIDE parameters improved the identification of fallers compared to the Timed Up and Go test.