Visualization and comparison of CT dose distribution between axial and helical acquisitions
Kai Yang1, Xinhua Li1, Lifeng Yu2
1Division of Diagnostic Imaging Physics, Department of Radiology, Massachusetts General Hospital, Boston, Massachusetts, USA.
Background:
It is challenging to assess the accuracy of volume CT Dose Index (CTDIvol ) when the axial scan modes corresponding to a helical scan protocol are not available. An alternative approach was proposed to directly measure C T D I v o l H $CTDI_{vol}^H$ using helical acquisitions and relatively small differences (< 20%) from CTDIvol were observed.
Purpose:
To visually demonstrate the 3D dose distribution for both axial and helical CT acquisitions and quantitively compare C T D I v o l H $CTDI_{vol}^H$ and CTDIvol .
Methods:
3D dose distribution within the standard CTDI phantoms (16 and 32 cm diameter) from a single CT projection, Dp (x,y,z) was first generated using Monte Carlo simulation (GEANT4) with 9×108 photons per combination of tube voltage (80-140 kV), collimation width (1-8 cm), and z-axis location of the central ray of the x-ray beam, with a spatial resolution of 1 mm3 . These dose distributions from one single projection were analytically ensembled to simulate 3D dose volumes DA (x,y,z) and DH (x,y,z) for axial and helical scans, respectively, with different helical pitches (0.3-2) and scan lengths (100-150 mm). 2D planar dose distributions were obtained by integrating the inside 100 mm of the dose volumes. CTDIvol and C T D I v o l H $CTDI_{vol}^H\;$ were calculated using the planar dose data at corresponding pencil chamber locations and the percentage differences (PD) were reported.
Results:
High spatial resolution 3D CT dose volumes were generated and visualized. PDs between C T D I v o l H $CTDI_{vol}^H$ and CTDIvol had strong dependency on scan length and peripheral chamber locations, with subtle dependency on collimation width and pitch. PDs were mostly within the range of ± 3% for a scan length of 150 mm with four peripheral chamber locations.
Conclusions:
With a scan length covering the entire phantom length, C T D I v o l H $CTDI_{vol}^H$ directly measured from helical scans can serve as an alternative to CTDIvol only if all four peripheral locations were measured.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT