Effect of multiple loading FIT sample collection devices on faecal haemoglobin concentration
Cerin John1, Carolyn Piggott1, Sally C Benton1,2
1NHS Bowel Cancer Screening Programme, Southern Hub, Guildford, UK.
Annals of Clinical Biochemistry
|June 7, 2023
Summary
Multiple sample loading into faecal collection devices significantly increases faecal haemoglobin concentration (f-Hb) measurements. This finding is crucial for accurate colorectal cancer (CRC) screening using quantitative faecal immunochemical tests (FIT).
Area of Science:
- Clinical Chemistry
- Gastroenterology
- Cancer Screening
Background:
- Faecal immunochemical tests (FIT) are vital for colorectal cancer (CRC) screening.
- Quantitative FIT aids in triaging symptomatic patients in primary care.
- Faecal sample collection devices (SCDs) with internal collars are used for sample collection.
Purpose of the Study:
- To investigate the impact of multiple sample loading on faecal haemoglobin (f-Hb) concentration.
- To assess this impact across four different FIT systems.
- To evaluate the effect of mixing versus no mixing between sample loads.
Main Methods:
- f-Hb negative samples were spiked with blood and homogenized.
- Samples were loaded 1, 3, and 5 times into SCDs, with and without mixing between loads.
- f-Hb was measured using the respective FIT systems, comparing multiple loads to a single load.
Main Results:
- Multiple loading (1-3 and 1-5 times) significantly altered f-Hb mass and concentration (p < 0.05) for all tested systems.
- The median percentage change in f-Hb was higher in the mixed sample group compared to the unmixed group.
- Significant differences were observed between mixed and unmixed groups for both 1-3 and 1-5 load scenarios.
Conclusions:
- Multiple loading into SCDs significantly increases f-Hb measurements.
- This effect impacts the accuracy of quantitative FIT results.
- Careful consideration of sample loading procedures is necessary for reliable CRC screening.