Related Experiment Video
Updated: Jul 27, 2025

Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
Magnetic Resonancy Imaging Safety in Active Osseointegrated (Osia®) and Cochlear Implants: New Technology Creating
Ibrahim Abdalla1, Janet S Choi2, Griffin Struyk1
1University of Minnesota School of Medicine, Minneapolis, Minnesota, U.S.A.
Objective:
There is increased confusion regarding MRI-compatible CIs and BAHAs. This report describes two cases when patients underwent MRIs with non-MRI compatible devices.
Results:
One patient with bilateral Cochlear Osias experienced dislocation of both internal magnets after 1.5 Tesla MRI. Both magnets were outside the silastic sheath, with the left magnet flipped. A second patient with a legacy CI experienced similar internal magnet dislocation and inversion after 3 Tesla MRI.
Conclusions:
This study describes internal magnet dislocation/inversion with the Cochlear Osia and a legacy CI after MRI. Our findings suggest the need for improved patient education and simplified radiology guidelines. Laryngoscope, 134:393-396, 2024.