Microperimetry as an Outcome Measure in RPGR-associated Retinitis Pigmentosa Clinical Trials

Laura J Taylor1,2, Amandeep S Josan1,2, Jasleen K Jolly3

  • 1Nuffield Laboratory of Ophthalmology, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, UK.

Abstract

Insights

Volume sensitivity is a superior microperimetry outcome measure for X-linked retinitis pigmentosa clinical trials. Pointwise sensitivity shows high variability, while volume sensitivity is unaffected by skewed data, making it more reliable for RPGR-associated RP.

Area of Science:

  • Ophthalmology
  • Genetics
  • Clinical Trials

Background:

  • X-linked retinitis pigmentosa (RP) is a genetic disorder.
  • RPGR gene mutations are a common cause of X-linked RP.
  • Microperimetry is used to assess visual function in retinal diseases.

Purpose of the Study:

  • To determine the most suitable microperimetry sensitivity index for clinical trials in X-linked RPGR-associated RP.
  • To compare the test-retest variability of pointwise, mean, and volume sensitivity measures.

Main Methods:

  • Retrospective analysis of microperimetry data from patients with RPGR-associated RP.
  • Repeatability analysis using triplicate testing over two days.
  • Longitudinal analysis of data from two separate visits.

Main Results:

  • Pointwise sensitivity showed high test-retest variability (CoR ±9.5 dB right, ±9.3 dB left).
  • Mean sensitivity had lower variability (CoR ±0.7 dB right, ±1.3 dB left) but was affected by skewed data.
  • Volume sensitivity demonstrated low variability (CoR ±144.5 dB*deg2 right, ±324.2 dB*deg2 left) and was unaffected by skewed data.

Conclusions:

  • Pointwise sensitivity indices are not recommended as outcome measures due to high variability.
  • Volume sensitivity indices are superior for RPGR-associated RP clinical trials compared to mean sensitivity.
  • Population-specific test-retest variability should be reported in clinical trials for accurate interpretation.