"Scan-(pre)Plan-Treat" Workflow for Bone Metastases Using the Ethos Therapy System: A Single-Center, In Silico
Eva Oldenburger1,2,3, Robin De Roover1,2, Kenneth Poels1,2
1Department of Radiation Oncology, University Hospitals Leuven, Leuven, Belgium.
Purpose:
To report on the accuracy of automated delineation, treatment plan quality, and duration of an in-silico "scan-(pre)plan-treat" (SPT) workflow for vertebral bone metastases using a 1 × 8 Gy regimen.
Method And Materials:
The cloud-based emulator system of the Ethos therapy system was used to adapt an organ-at-risk-sparing preplan created on the diagnostic CT to the anatomy-of-the-day using the cone beam CT made before treatment.
Results:
SPT using the Ethos emulator system resulted in relatively good coverage of the PTV and acceptable dose to the OAR. Delivery time and plan homogeneity was the best for 7-field IMRT plan template.
Conclusions:
A SPT workflow formula results in a highly conformal treatment delivery while maintaining an acceptable timeframe for the patient on the treatment couch.