Comparison of Low-Level to High-Level Disinfection in Eliminating Microorganisms From Ultrasound Transducers Used on
Nathan Peters1,2,3, Frances Williamson2,4, Michelle J Bauer5
1Department of Anaesthesia and Perioperative Medicine, Royal Brisbane and Women's Hospital, Brisbane, Australia.
Introduction:
There is a lack of international consensus as to whether high- or low-level disinfection (HLD or LLD) is required for ultrasound (US) transducers used during percutaneous procedures. This study compared the effectiveness of LLD to HLD on US transducers contaminated with microorganisms from skin.
Methods:
Two identical linear US transducers repeatedly underwent either LLD or HLD during the study. Randomization determined which of these transducers was applied to left and right forearms of each participant. Swabs taken from transducers before and after reprocessing were plated then incubated for 4-5 days, after which colony forming units (CFU) were counted and identified. The primary hypothesis was the difference in the proportion of US transducers having no CFUs remaining after LLD and HLD would be less than or equal to the noninferiority margin of -5%.
Results:
Of the 654 recruited participants 73% (n = 478) had microbial growth from both transducers applied to their left and right forearms before reprocessing. These were included in the paired noninferiority statistical analysis where, after disinfection, all CFUs were eliminated in 100% (95% CI: 99.4-100.0%) of HLD transducer samples (n = 478) and 99.0% (95% CI: 97.6-99.7%) of LLD transducer samples (n = 473). The paired difference in the proportion of transducers having all CFUs eliminated between LLD and HLD was -1.0% (95% CI: -2.4 to -0.2%, P-value <.001).
Conclusions:
Disinfection with LLD is noninferior to HLD when microorganisms from skin have contaminated the transducer. Therefore, using LLD for US transducers involved in percutaneous procedures would present no higher infection risk compared with HLD.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Hand hygiene
Hand washing...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Antimicrobial Effectiveness
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...