Purpose:
Optimal size selection is important for successful Woven EndoBridge (WEB) treatment. Conventional recommendations for WEB sizing based on aneurysm width and height sometimes require device exchange. We aimed to design a novel volume-based parameter, the ideal WEB-aneurysm volume (iWAVe) ratio, for optimal WEB sizing.
Methods:
Consecutive patients who underwent WEB treatment for wide-neck bifurcation aneurysms between January 2021 and May 2022 were retrospectively reviewed. Aneurysm volume was automatically calculated using software. We measured the aneurysm volume based on the expected position of the device within the aneurysm. The WAVe ratio was defined as the ratio of the aneurysm volume to WEB volume. We dichotomized aneurysms treated with a successful sizing or unsuccessful sizing for WEB (successful group and unsuccessful group, respectively).
Results:
Thirty-five patients were eligible for study enrollment. Ten patients (28.6%) needed to exchange the WEB on the first attempt and required another WEB on the second attempt resulting in deployment success. Hence, 35 aneurysms were in the successful group and 10 were in the unsuccessful group. The median WAVe ratio was 1.0 (range 0.76-1.31) in the successful group and 1.27 (0.58-1.89) in the unsuccessful group. Using logistic regression, iWAVe ratio was from 0.90-1.16 to secure a >80% probability of success by the 95% lower confidence limit. The sensitivity and specificity of the iWAVe ratio for optimal size selection on the first attempt were 0.60 and 1.00, respectively.
Conclusion: