Evaluation of Antimicrobial Susceptibility Testing Performance of DL 96 System for Enterobacterales
Hua Wu1, Xiaojun Zhou1, Fan Zhou2
1Department of Laboratory Medicine, Hainan General Hospital, Affiliated Hainan Hospital of Hainan Medical College, Haikou, People's Republic of China.
Abstract:
DL96 Microbial Identification/Antimicrobial Susceptibility Testing (ID/AST) System (Zhuhai DL, Guangdong, China) is one of the most commonly used commercial ID/AST System in China. This study aims to evaluate the performance of DL 96E for Antimicrobial Susceptibility Testing (AST) of 270 Enterobacterales isolates from Hainan general hospital using the broth microdilution method (BMD) as reference method. CLSI M52 criteria was followed when analyzing the evaluation results. Twenty antimicrobial agents were evaluated, and categorical agreement (CA) ranged from 62.8% to 96.5%. Imipenem had the lowest CA (63.9%) and highest very major errors (VME) (52.8%). A total of 103 carbapenem-resistant Enterobacterales were evaluated; DL 96E miss identified 22 isolates, including six carbapenemase-producing Enterobacteriaceae. DL 96E must adjust the Minimum Inhibitory Concentration (MIC) ranges of ciprofloxacin, levofloxacin, and piperacillin-tazobactam to cover Clinical and Laboratory Standards Institute (CLSI) breakpoints, adjust the formulation of some antimicrobial, such as imipenem, and increase the MIC detection range to cover the Quality control (QC) strains' MIC range.
More Related Videos
12:03Antimicrobial Synergy Testing by the Inkjet Printer-assisted Automated Checkerboard Array and the Manual Time-kill Method
Published on: April 18, 2019
12:08Rapid Antimicrobial Susceptibility Testing by Stimulated Raman Scattering Imaging of Deuterium Incorporation in a Single Bacterium
Published on: February 14, 2022