Initial evaluation of accelerator-based neutron source system at the Shonan Kamakura General Hospital
Shunsuke Suzuki1, Kazunori Nitta2, Takayuki Yagihashi2
1Department of Medical Physics, Shonan Kamakura General Hospital, 1370-1 Okamoto, Kamakura, Kanagawa, 247-8533, Japan; Graduate School of Engineering, Kyoto University, Kyoto University Katsura, Kyoto Nishikyo-ku, Kyoto, 615-8246, Japan.
Abstract:
An accelerator-based boron neutron capture therapy (AB-BNCT) system was installed at the Shonan Kamakura General Hospital (SKGH). We confirmed that a stable operation was possible for 1 h at a current of 30 mA. The evaluated thermal neutron flux was 2.8 × 109 cm-2 s-1 and in good agreement (±5%) with the calculated values. The daily variation was within ±2%. The ambient dose rate due to residual radioactivity after irradiation was approximately 5 μSv/h using a lead shutter.