Photon-counting CT for diagnosis of acute pulmonary embolism: potential for contrast medium and radiation dose reduction
Related Experiment Video
Updated: Jul 26, 2025

High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Pauline Pannenbecker1, Henner Huflage2, Jan-Peter Grunz2
1Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacherstr. 6, D-97080, Würzburg, Germany. pannenbeck_p@ukw.de.
Objective:
To evaluate the image quality of an ultra-low contrast medium and radiation dose CT pulmonary angiography (CTPA) protocol for the diagnosis of acute pulmonary embolism using a clinical photon-counting detector (PCD) CT system and compare its performance to a dual-energy-(DE)-CTPA protocol on a conventional energy-integrating detector (EID) CT system.
Methods:
Sixty-four patients either underwent CTPA with the novel scan protocol on the PCD-CT scanner (32 patients, 25 mL, CTDIvol 2.5 mGy·cm) or conventional DE-CTPA on a third-generation dual-source EID-CT (32 patients, 50 mL, CTDIvol 5.1 mGy·cm). Pulmonary artery CT attenuation, signal-to-noise ratio, and contrast-to-noise-ratio were assessed as objective criteria of image quality, while subjective ratings of four radiologists were compared at 60 keV using virtual monoenergetic imaging and polychromatic standard reconstructions. Interrater reliability was determined by means of the intraclass correlation coefficient (ICC). Effective dose was compared between patient cohorts.
Results:
Subjective image quality was deemed superior by all four reviewers for 60-keV PCD scans (excellent or good ratings in 93.8% of PCD vs. 84.4% of 60 keV EID scans, ICC = 0.72). No examinations on either system were considered "non-diagnostic." Objective image quality parameters were significantly higher in the EID group (mostly p < 0.001), both in the polychromatic reconstructions and at 60 keV. The ED (1.4 vs. 3.3 mSv) was significantly lower in the PCD cohort (p < 0.001).
Conclusions:
PCD-CTPA allows for considerable reduction of contrast medium and radiation dose in the diagnosis of acute pulmonary embolism, while maintaining good to excellent image quality compared to conventional EID-CTPA.
Clinical Relevance Statement:
Clinical PCD-CT allows for spectral assessment of pulmonary vasculature with high scan speed, which is beneficial in patients with suspected pulmonary embolism, frequently presenting with dyspnea. Simultaneously PCD-CT enables substantial reduction of contrast medium and radiation dose.
Key Points:
• The clinical photon-counting detector CT scanner used in this study allows for high-pitch multi-energy acquisitions. • Photon-counting computed tomography allows for considerable reduction of contrast medium and radiation dose in the diagnosis of acute pulmonary embolism. • Subjective image quality was rated best for 60-keV photon-counting scans.
More Related Videos
13:10Combined Near-infrared Fluorescent Imaging and Micro-computed Tomography for Directly Visualizing Cerebral Thromboemboli
Published on: September 25, 2016
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
Related Concept Videos
Pulmonary Embolism II: Diagnostic Studies and Interprofessional Care
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...