Investigation on the proton range uncertainty with spectral CT-based virtual monoenergetic images
Libing Zhu1, Yi Du2, Yahui Peng3
1Institute of Nuclear and New Energy Technology, Tsinghua University, Beijing, PR China.
Objective:
The stopping power ratio (SPR) prediction error will contribute to the range uncertainty of proton therapy. Spectral CT is promising in reducing the uncertainty in SPR estimation. The purpose of this research is to determine the optimal energy pairs of SPR prediction for each tissue and to evaluate the dose distribution and range difference between the spectral CT with the optimal energy pairs method and the single energy CT (SECT) method.
Methods:
A new method was proposed based on image segmentation to calculate the proton dose with spectral CT images for the head and body phantom. CT number of each organ region were converted to SPR with the optimal energy pairs of each organ. The CT images were segmented into different organ parts with thresholding method. Virtual monoenergetic (VM) images from 70 keV to 140 keV were investigated to determine the optimal energy pairs for each organ based on Gammex 1467 phantom. The beam data of Shanghai Advanced Proton Therapy facility (SAPT) was employed in matRad (an open-source software for radiation treatment planning) for the dose calculation.
Results:
The optimal energy pairs were obtained for each tissue. The dose distribution of two tumor sites (brain and lung) were calculated with the aforementioned optimal energy pairs. The maximum dose deviation between spectral CT and SECT at the target region was 2.57% and 0.84% for the lung tumor and brain tumor respectively. The range difference between spectral and SECT was significant with 1.8411 mm for the lung tumor. γ passing rate was 85.95% and 95.49% for the lung tumor and brain tumor with the criterion 2%/2 mm.
Conclusions:
This work presents a way to determine the optimal energy pairs for each organ and to calculate the dose distribution based on the more accurate SPR prediction.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
08:46Using Synchrotron Radiation Microtomography to Investigate Multi-scale Three-dimensional Microelectronic Packages
Published on: April 13, 2016
Related Concept Videos
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...