Low-Cost SpO2 Integrated Neonatal CPAP Device for Low Resource Setting
Ahmed Ali Dawud1, Ahmed Mohammed Abagaro1
1School of Biomedical Engineering, Jimma Institute of Technology, Jimma University, Jimma, Ethiopia.
Introduction:
More than 60% of preterm births take place in South Asia and sub-Saharan Africa, making prematurity a primary cause of neonatal mortality. Even though continuous positive airway pressure (CPAP) is a popular treatment for respiratory distress syndrome (RDS) and is safe, practicable, and efficient for use in LMICs, it is crucial to ensure that neonates receive the full benefits of the therapy by monitoring their blood oxygen level.
Methods:
A centrifugal fan, power source, control system, and sensors are all included in our design. A centrifugal fan was created to provide air at positive pressure in the range of approximately 4 cmH2O to 20 cmH2O utilizing revolving blades (impeller), a DC motor, and a fixed component. The control unit contains a microcontroller to handle sensor data. The proportional-integral (PI) controller board's external potentiometer is used to set the pressure level.
Results:
To ascertain whether the prototype satisfies the design requirements, it was constructed and put through several iterations and testing. The proposed device's prototype was tested for accuracy, affordability, and usability. The centrifugal fan speed measurement was accurate to within 94.5%, while the oxygen concentration sensor reading was accurate to within 98.5%.
Conclusion:
The design investigates viability of a straightforward, inexpensive, portable SpO2 integrated neonatal CPAP device for use in the delivery room in low-resource countries and to evaluates methods for measuring flows during CPAP treatment by monitoring the level of oxygen in the blood and pressure level delivered by the device using the lowest and safest setting that yields useful results.
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: