Variability in PET image quality and quantification measured with a permanently filled 68Ge-phantom: a multi-center
O Sipilä1, J Liukkonen2, H-L Halme3
1HUS Diagnostic Center, Clinical Physiology and Nuclear Medicine, Helsinki University Hospital and University of Helsinki, P. O. Box 442, 00029, Helsinki, Finland. outi.sipila@hus.fi.
Background:
This study evaluated, as a snapshot, the variability in quantification and image quality (IQ) of the clinically utilized PET [18F]FDG whole-body protocols in Finland using a NEMA/IEC IQ phantom permanently filled with 68Ge.
Methods:
The phantom was imaged on 14 PET-CT scanners, including a variety of models from two major vendors. The variability of the recovery coefficients (RCmax, RCmean and RCpeak) of the hot spheres as well as percent background variability (PBV), coefficient of variation of the background (COVBG) and accuracy of corrections (AOC) were studied using images from clinical and standardized protocols with 20 repeated measurements. The ranges of the RCs were also compared to the limits of the EARL 18F standards 2 accreditation (EARL2). The impact of image noise on these parameters was studied using averaged images (AVIs).
Results:
The largest variability in RC values of the routine protocols was found for the RCmax with a range of 68% and with 10% intra-scanner variability, decreasing to 36% when excluding protocols with suspected cross-calibration failure or without point-spread-function (PSF) correction. The RC ranges of individual hot spheres in routine or standardized protocols or AVIs fulfilled the EARL2 ranges with two minor exceptions, but fulfilling the exact EARL2 limits for all hot spheres was variable. RCpeak was less dependent on averaging and reconstruction parameters than RCmax and RCmean. The PBV, COVBG and AOC varied between 2.3-11.8%, 9.6-17.8% and 4.8-32.0%, respectively, for the routine protocols. The RC ranges, PBV and COVBG were decreased when using AVIs. With AOC, when excluding routine protocols without PSF correction, the maximum value dropped to 15.5%.
Conclusion:
The maximum variability of the RC values for the [18F]FDG whole-body protocols was about 60%. The RC ranges of properly cross-calibrated scanners with PSF correction fitted to the EARL2 RC ranges for individual sphere sizes, but fulfilling the exact RC limits would have needed further optimization. RCpeak was the most robust RC measure. Besides COVBG, also RCs and PVB were sensitive to image noise.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
10:02Whole-body PET/MRI of Pediatric Patients: The Details That Matter
Published on: December 19, 2017
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...