Trabecular Meshwork Response to Pressure Elevation in the Living Human Eye
Published on: June 20, 2015
The Repeatability and Reproducibility of a New Minimally Invasive Method for Orbital Pressure Measurement
Xue Jiang1, Li Dong, Minghui Wang
1Beijing Tongren Eye Center, Beijing Ophthalmology and Visual Sciences Key Lab, Capital Medical University, Beijing Tongren Hospital, Beijing, China.
Background:
Evaluation of orbital pressure is crucial for monitoring various orbital disorders. However, there is currently no reliable technique to accurately measure direct orbital pressure (DOP). This study aimed to establish a new method for the DOP as well as to verify its repeatability and reproducibility in rabbits.
Methods:
The study included 30 normal eyes from fifteen 3-month-old New Zealand white rabbits. After administering inhalation anesthesia, intraocular pressure (IOP) was determined by tonometry (Tonopen). For DOP manometry, a TSD104 pressure transducer was inserted between the disposable injection needle and the syringe, and the output results were displayed on a computer. Two observers independently participated in the experiment to verify its repeatability and reproducibility.
Results:
The mean IOP of rabbits was significantly higher than the DOP in normal rabbits (11.67 ± 1.08 mm Hg versus 4.91 ± 0.86 mm Hg, P < 0.001). No significant interocular difference was detected for both IOP and DOP ( P > 0.05). A high correlation was found for intraobserver measurements of both IOP (intraclass correlation coefficient = 0.87, P < 0.001) and DOP (intraclass correlation coefficient = 0.89, P < 0.001). A high agreement was also presented for the interobserver reproducibility for the measurements of IOP [Pearson correlation coefficient ( R ) = 0.86, P < 0.001] and DOP ( R = 0.87, P < 0.001). Direct orbital pressure was positively correlated with IOP in both observers ( R 1 = 0.66, R 2 = 0.62, P < 0.001). The Bland-Altman plots revealed that 5.0% (3/60) of the IOP and DOP measurement points were outside of the 95% limits of agreement, respectively.
Conclusions:
The TSD104 pressure transducer-based manometry may serve as a reliable device for the measurement of DOP, providing real-time measuring results with acceptable reproducibility and repeatability.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...