Analytical verification of the Dymind D7-CRP automated analyser
Merima Čolić1, Bojana Magdić1, Monika Kolundžić1
1Department of Laboratory Diagnostics, Children's Hospital Zagreb, Zagreb, Croatia.
Introduction:
The aim of this study was to perform a verification of the Dymind D7-CRP automated analyser and compare it with established analysers.
Materials And Methods:
Analytical verification included estimation of repeatability, between run precision, within-laboratory precision, and bias in control samples with low, normal and high levels. The acceptance criteria for analytical verification were defined using the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) 2019 Biological Variation Database. Method comparison between the Dymind D7-CRP and the Sysmex XN1000 for haematological parameters and the Dymind D7-CRP and the Beckman Coulter AU680 for CRP values was performed on 40 patient samples.
Results:
Analytical verification criteria were adequately met with the exception of monocyte count for repeatability and within-laboratory precision (13.4% and 11.5%, respectively, acceptance criteria 10.1%) and measurement uncertainty (23.0, acceptance criteria 20.0%) at low level, eosinophil count for BIAS at the low level (37.7%, acceptance criteria 25.2%), basophil count (BAS) for BIAS at the high level (14.2%, acceptance criteria 10.9%), and mean platelet volume (MPV) for repeatability (4.2% and 6.8%), between run precision (2.2% and 4.7%), within-laboratory precision (4.0% and 7.3%) (acceptance criteria 1.7%), and measurement uncertainty (8.0 and 14.6%, acceptance criteria 3.4%) at both the low and high concentrations. Method comparison showed no clinically significant constant or proportional differences for all parameters except BAS and MPV.
Conclusion:
The analytical verification of the Dymind D7-CRP showed adequate analytical characteristics. The Dymind D7-CRP can be used interchangeably with the Sysmex XN-1000 for all parameters tested, except BAS and MPV, and with the Beckman Coulter AU-680 for the determination of CRP.