CD Spectroscopy to Study DNA-Protein Interactions
Published on: February 10, 2022
Can we still measure circular dichroism with circular dichroism spectrometers: The dangers of anisotropic artifacts
Thomas J Ugras1,2, Yuan Yao3, Richard D Robinson2,3
1School of Applied and Engineering Physics, Cornell University, Ithaca, New York, USA.
Abstract:
Chiral materials with strong linear anisotropies are difficult to accurately characterize with circular dichroism (CD) because of artifactual contributions to their spectra from linear dichroism (LD) and birefringence (LB). Historically, researchers have used a second-order Taylor series expansion on the Mueller matrix to model the LDLB interaction effects on the spectra in conventional materials, but this approach may no longer be sufficient to account for the artifactual CD signals in emergent materials. In this work, we present an expression to model the measured CD using a third-order expansion, which introduces "pairwise interference" terms that, unlike the LDLB terms, cannot be averaged out of the signal. We find that the third-order pairwise interference terms can make noticeable contributions to the simulated CD spectra. Using numerical simulations of the measured CD across a broad range of linear and chiral anisotropy parameters, the LDLB interactions are most prominent in samples that have strong linear anisotropies (LD, LB) but negligible chiral anisotropies, where the measured CD strays from the chirality-induced CD by factors greater than 103 . Additionally, the pairwise interactions are most significant in systems with moderate-to-strong chiral and linear anisotropies, where the measured CD is inflated twofold, a figure that grows as linear anisotropies approach their maximum. In summary, media with moderate-to-strong linear anisotropy are in great danger of having their CD altered by these effects in subtle manners. This work highlights the significance of considering distortions in CD measurements through higher-order pairwise interference effects in highly anisotropic nanomaterials.
More Related Videos
Related Concept Videos
Spectrophotometry: Introduction
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
IR Spectrometers
UV–Vis Spectrometers
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
UV–Vis Spectroscopy: Woodward–Fieser Rules
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...