Establishing a quality assurance program for photon counting detector (PCD) CT: Tips and caveats
Zaki Ahmed1,2, Andrea Ferrero1, Liqiang Ren1,3
1Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
A quality assurance program for photon-counting-detector CT (PCD-CT) using the ACR CT manual is suitable for accreditation, but proper parameter selection is crucial. Additional evaluations are recommended for comprehensive PCD-CT performance assessment.
Area of Science:
- Medical Imaging Physics
- Radiological Sciences
- Diagnostic Imaging Technology
Background:
- Photon-counting-detector CT (PCD-CT) offers unique capabilities beyond conventional CT systems.
- Evaluating PCD-CT performance requires specialized quality assurance (QA) protocols.
- The American College of Radiology (ACR) CT quality control (QC) manual provides a framework for CT system evaluation.
Purpose of the Study:
- To assess the adequacy of an ACR CT QC manual-based QA program for evaluating a clinical PCD-CT system.
- To determine if the existing QA program can fully assess the advanced features of PCD-CT technology.
Main Methods:
- Implemented a daily QA program assessing CT number accuracy and artifacts in standard and ultra-high-resolution (UHR) modes.
- Conducted a full system evaluation using the ACR CT Accreditation Phantom and routine protocols.
- Reconstructed images at various keV levels and evaluated spatial resolution via modulation transfer function (MTF) and iodine insert accuracy.
Main Results:
- The QA program successfully identified detector issues requiring recalibration or replacement.
- CT number accuracy varied by image type; 70 keV virtual monoenergetic images (VMIs) met ACR criteria, while others did not.
- Spatial resolution exceeded ACR phantom limits (~40 lp/cm vs. 12 lp/cm), and iodine quantification was accurate across VMIs.
Conclusions:
- Careful selection of protocols and parameters on PCD-CT systems is essential for meeting ACR accreditation standards.
- Utilizing the 70 keV VMI enabled the system to pass all ACR CT manual tests.
- Comprehensive PCD-CT evaluation necessitates supplementary tests like MTF measurements and multi-energy phantom scans.