Assessing target localization accuracy across different soft-tissue matching protocols using end-exhalation
Makoto Sasaki1, Mitsuhiro Nakamura2,3, Ryo Ashida3
1Division of Clinical Radiology Service, Kyoto University Hospital, 54 Kawaharacho, Shogoin, Sakyo-ku, Kyoto 606-8507, Japan.
Abstract:
The purpose of this study was to retrospectively assess target localization accuracy across different soft-tissue matching protocols using cone-beam computed tomography (CBCT) in a large sample of patients with pancreatic cancer and to estimate the optimal margin size for each protocol. Fifty-four consecutive patients with pancreatic cancer who underwent 15-fraction volumetric modulated arc therapy under the end-exhalation breath-hold condition were enrolled. Two soft-tissue matching protocols were used according to the resectability classification, including gross tumor volume (GTV) matching for potentially resectable tumors and planning target volume (PTV) matching for borderline resectable or unresectable tumors. The tolerance of the target localization error in both matching protocols was set to 5 mm in any direction. The optimal margin size for each soft-tissue matching protocol was calculated from the systematic and random errors of the inter- and intrafraction positional variations using the van Herk formula. The inter- and intrafraction positional variations of PTV matching were smaller than those of GTV matching. The percentage of target localization errors exceeding 5 mm in the first CBCT scan of each fraction in the superior-inferior direction was 12.6 and 4.8% for GTV and PTV matching, respectively. The optimal margin sizes for GTV and PTV matching were 3.7 and 2.7, 5.4 and 4.1 and 3.9 and 3.0 mm in the anterior-posterior, superior-inferior and left-right directions, respectively. Target localization accuracy in PTV matching was higher than that in GTV matching. By setting the tolerance of the target localization error, treatment can be successful within the planned margin size.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
08:17Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015