Related Experiment Video
Updated: Sep 13, 2026

Electrochemiluminescence Assays for Human Islet Autoantibodies
Published on: March 23, 2018
Establishment and analytical performance of light-initiated chemiluminescence assay method for detecting thyrotropin
Ying Sun1, Mengxue Su2, Miao Wang3
1Department of Clinical Laboratory, Tianjin Medical University General Hospital, Tianjin, 300052, People's Republic of China.
Abstract:
Aim: To evaluate double-antibody competitive light-initiated chemiluminescence assay method for detecting the thyrotropin receptor antibody. Materials & methods: The optimal working concentrations of competitive antibody and rTSHR were confirmed by checkerboard titration. Assay performance was assessed by precision, linearity, accuracy, limit of blank and clinical evaluation. Results: The coefficient of variation for repeatability and intermediate precision was 3.9-5.9 and 0.9-1.3%, respectively. The correlation coefficient was 0.999 by least squares linear fitting in linearity evaluation. The relative deviation ranged from -5.9 to 4.1%, and the limit of blank of the method was 0.13 IU/l. Compared with the Roche cobas system (Roche Diagnostics, Mannheim, Germany), the relationship between the two assays was shown to be significantly correlative. Conclusion: The light-initiated chemiluminescence assay method for detecting thyrotropin receptor antibody is a rapid, novel and accurate method for thyrotropin receptor antibody measurement.

