Related Experiment Video
Updated: Jul 25, 2026

Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Deep Inspiratory Breath-Hold Radiation for Left-Sided Breast Cancer using Novel Frame-based Tactile Feedback
Sapna Nangia1,2, Robin Khosa1, Divya Piyushi1,3
1Department of Radiation Oncology, Indraprastha Apollo Hospital, Chennai, Tamil Nadu, India.
Abstract:
A frame providing tactile feedback for the reproducibility of deep inspiratory breath-hold (DIBH) is described. The frame, fitted across the patient, comprises a horizontal bar, parallel to the patient's long axis, and holds a graduated pointer perpendicular to it. The pointer provides individualized tactile feedback for reproducibility of DIBH. Within the pointer is a movable pencil, bearing a 5 mm coloured strip which becomes visible only during DIBH, and acts as a visual cue to the therapist. The average variation in separation in the planning and pretreatment cone-beam computed tomography of 10 patients was 2 mm (confidence interval 1.95-2.05). Frame-based tactile feedback is a novel, reproducible technique for DIBH.