Purpose:
We have been using a paper-based hard copy print (paper print) system of X-ray images, in which digital imaging and communications in medicine (DICOM) data can be directly output on papers from medical imaging systems or from a picture archiving and communication system (PACS) server, and they are utilized as patient referral materials or for preoperative planning. The purpose of this study was to compare the display performance of X-ray images on the printed paper and that on the liquid crystal display (LCD).
Methods:
We measured contrast response to verify consistency of image appearance on both display systems. The contrast resolution was assessed by a CDRAD phantom. The spatial resolution was assessed by an X-ray test chart.
Results:
The contrast response of the paper printer was not concordant with the grayscale standard display function (GSDF). The difference between the measured contrast response and the ideal GSDF on the paper was large in the high-density area. The low-contrast resolution on the paper was inferior to that on the LCD. The spatial resolving power on the paper was superior to that on the LCD.
Conclusion:
The display performance of the paper printer for X-ray images was clarified. X-ray images printed on the paper should be used carefully taking account of their characteristics of display performance.
More Related Videos
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...