Related Experiment Video
Updated: Jul 25, 2026

Low-Cost, Volume-Controlled Dipstick Urinalysis for Home-Testing
Published on: May 8, 2021
Analytical evaluation and quality assessment of the ARKRAY ADAMS A1c Lite HA-8380V for HbA1c
1Laboratory, Hospital Galdakao - Usansolo, Galdakao, Vizcaya, Spain.
Introduction:
ARKRAY ADAMS A1c Lite HA-8380V is a fully automated high- performance liquid chromatography system for the measurement of HbA1c. We aimed to evaluate its analytical performance using the Variant Mode.
Methods:
Carry-over, linearity, imprecision, trueness, interferences and comparison against ARKRAY ADAMS A1c HA-8180V and HA-8180T analyzers were studied.
Results:
Total CVs 0.93% (IFCC units), 0.63% (NGSP units) at low concentration and 1.01% (IFCC) 0.74% (NGSP), at high concentration. Mean difference with the target values was -0.44 mmol/mol (IFCC) -0.04% (NGSP). Carry-over, linearity and method comparison were excellent.The results were not affected in the range of total Hb 39-199 g/L, labile fraction 5.7%, carbamylated Hb 9.1% nor acetylated Hb 7.8%, effect of common variants was negligible.
Conclusions:
the analyzer demonstrated very good analytical performances, according to the consensus criteria established for HbA1c; it is adequate for laboratories with medium-low workloads.
More Related Videos
05:58Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
07:22Glycemic Impact on Knee Osteoarthritis Symptoms on Physical, Radiographic, and Inflammatory Markers among Individuals Aged 50 and Over with Diabetes
Published on: March 7, 2025
Related Concept Videos
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)