Low complexity adaptive ultrasound image beamformer combined with improved multiphase apodization with
Ping Wang1, Jinghan Chen1, Yue Shen1
1State Key Laboratory of Power Transmission Equipment and System Security and New Technology, Chongqing University, Chongqing 400044, China.
Abstract:
In this paper, an ultrasound imaging method combined with low-complexity adaptive beamformer (LCA) and improved multiphase apodization with cross-correlation (IMPAX) is proposed to improve image resolution and contrast with low hardware cost. Firstly, the delayed echo signal is apodized by the LCA to obtain a narrow mainlobe width echo signal and LCA output. Then, multiple pairs of complementary square-wave phase apodizations are applied to the apodized echo signal to obtain corresponding signal pairs, which are used to calculate the normalized cross-correlation (NCC) matrix. Finally, the average value of the NCC matrices is filtered by 2-D means, and the filtered result is introduced as the weighting factor for the LCA output. The simulation and experimental results show that the proposed LCA-IMPAX can effectively reduce the mainlobe width, suppress clutter, and be robust to noise. Compared with DAS, LCA, and MPAX, for simulated point targets, the full-width at half-maximum (FWHM, -6dB) of LCA-IMPAX is reduced by 49.22%, 10.06%, and 48.67%, respectively. For simulated cyst, the CR is improved by 219.91%, 138.08%, and 103.44%, respectively. For experimental cysts, the CR is improved by an average of 145.00%, 136.14%, and 55.09%, respectively. The results of human heart data indicate that LCA-IMPAX has good imaging quality in vivo. Since the proposed method does not involve covariance matrix inversion, it can be applied in real-time imaging systems.