Validation of Segmental Post-PCI Physiological Gradients With IVUS-Detected Focal Lesions and Stent Underexpansion
Tara Neleman1, Alessandra Scoccia1, Frederik T W Groenland1
1Department of Cardiology, Erasmus University Medical Center, Rotterdam, the Netherlands.
Background:
Segmental postpercutaneous coronary intervention (PCI) pressure gradients may detect residual disease and potential targets for optimization. However, universal definitions of relevant segmental gradients are lacking.
Objectives:
This study sought to evaluate the diagnostic performance of post-PCI fractional flow reserve (FFR), distal coronary pressure-to-aortic pressure ratio (Pd/Pa), and diastolic pressure ratio (dPR) gradients to detect residual focal lesions and stent underexpansion as observed by intravascular ultrasound (IVUS).
Methods:
Patients from the IVUS-guided optimization arm of the FFR REACT (FFR-guided PCI Optimization Directed by High-Definition IVUS Versus Standard of Care) trial with complete IVUS and FFR pullback data were included. Patients with angiographically successful PCI and post-PCI FFR <0.90 underwent FFR, Pd/Pa, and IVUS pullbacks. dPR was calculated offline using dedicated software. Segmental pressure gradients (distal, in stent, and proximal) in segments ≥5 mm were evaluated against IVUS-detected residual disease (distal or proximal focal lesions and stent underexpansion).
Results:
A total of 139 vessels were included (mean post-PCI FFR: 0.83 ± 0.05, range 0.56-0.89). Focal distal and proximal lesions were detected by IVUS in 23 (17.4%) of 132 and 14 (12.6%) of 111 vessels, respectively, whereas stent underexpansion was present in 86 (61.9%) vessels. Diagnostic ability of segmental FFR gradients to predict IVUS-detected distal and proximal lesions was moderate-to-good (area under the curve [AUC]: 0.69 and 0.84, respectively) and poor to moderate for segmental Pd/Pa and dPR gradients (AUC ranging from 0.58 to 0.69). In-stent gradients had no discriminative ability to detect stent underexpansion (FFR AUC: 0.52; Pd/Pa AUC: 0.54; dPR AUC: 0.55).
Conclusions:
In patients with post-PCI FFR <0.90, segmental post-PCI pressure gradients have moderate discriminative ability to identify IVUS-detected focal lesions but no discriminative ability to identify IVUS-detected stent underexpansion.