Precision and trueness verification study of an Atellica system
Alberto Vílchez Rodríguez1, Julia González Cantó1, Sara Esteve Poblador1
1Area of Biological Diagnosis, La Ribera University Hospital, Alzira, Spain.
Objectives:
Clinical laboratories should use only validated procedures. Precision is an important factor in the validation and verification of a new measurement procedure. Our objective was to verify the precision and trueness of different analysers used for the biochemical and immunochemical characterization of analytes.
Methods:
Advia 1800®, Immulite®2000 and CentaurXP® analysers and the Atellica®Solution system were used. Five analytes were characterized biochemically, whereas another five analytes were characterized immunochemically. Imprecision was assessed using BioRad® and Siemens® control materials. Within-run and between-run imprecision were calculated by analysing three replicates of each control in a single run every day for five days. Bias was assessed using 40 samples of serum by the analysis of differences and linear regression.
Results:
The within-run and between-run imprecision values obtained with the new measurement procedure were lower than the ones claimed by the manufacturer for all the analytes studied. In the bias study, a proportional but not constant systematic error was observed in some analytes.
Conclusions:
The coefficients of variation obtained with Atellica®Solution verified both, the imprecision specifications claimed by the manufacturer and by the laboratory. The conditions of calibration should be revised for some parameters and a wider range of samples should be used.
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Accuracy, limits, and approximation
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...
Improving Translational Accuracy
Quality Assurance
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Random and Systematic Errors