Image Contrast and Spectral Transmission in Intraocular Lenses with Nd:YAG Pits
A F Borkenstein1, E M Borkenstein1, P Omidi2
1Borkenstein & Borkenstein Private Practice, Privatklinik der Kreuzschwestern Graz, Graz, Austria.
Purpose:
Neodymium yttrium aluminum garnet (Nd:YAG) laser capsulotomy is considered as safe and effective method in the treatment of posterior capsule opacification. Nevertheless, side effects are described. The incorrectly adjusted focus of the laser beam during the procedure can lead to so-called YAG-pits or YAG-shots. In this experimental study, we measured spectral transmission to evaluate the image contrast and analyze the impact of YAG-pits in intraocular lenses (IOL).
Methods:
Acrylic, foldable, one-piece IOLs with 6.0 mm optic and different material properties were studied. These included: monofocal IOLs and enhanced monofocal IOLs with water content of 0.3%, 26.0%, and 4.0% and a refractive index of 1.49, 1.46, and 1.54, respectively. All measurements were done with new, unaltered IOLs and IOLs with YAG-pits. Damage was intentionally created, performing YAG-pits (n = 7) in the central zone (3.5 mm) using a photodisruption laser (2.0mJ). All laboratory measurements were repeated: These included surface topography characterization, United States Air Force (USAF) resolution test chart analysis, spectral transmittance measurements and through focus contrast measurement.
Results:
Significant differences were found between the unaltered lenses and lenses with defects. The YAG-pits within the optic of the IOLs decreased the image contrast and spectral transmission and changed results of USAF test images at the focal position by 62%, 57% and 54%, respectively. In all IOLs a reduction of the relative intensity of total transmitted light was observed between 450 and 700 nm wavelength.
Conclusion:
This experimental study confirmed that the IOL image performance deteriorates with YAG-pits. The total intensity of transmitted light or transmittance (without scattering) was reduced in the wavelength between 450 and 700 nm. The contrast was significantly reduced and USAF test targets showed much worse results compared to unmodified counterparts. There was no systematic difference between monofocal and enhanced monofocal lenses. Further experiments should investigate the effect of YAG-pits on diffractive IOLs.
More Related Videos
Related Concept Videos
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Glaucoma: Overview
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...