Evaluating Viscosity and Tear Breakup Time of Contemporary Commercial Ocular Lubricants on an In Vitro Eye Model
Chau-Minh Phan1,2, Mitchell Ross3, Karim Fahmy4
1Centre for Ocular Research & Education (CORE), School of Optometry and Vision Science, University of Waterloo, Waterloo, ON, Canada.
Purpose:
To evaluate the link between the viscosity of ophthalmic formulation and tear film stability using a novel in vitro eye model.
Methods:
The viscosities and noninvasive tear breakup time (NIKBUT) of 13 commercial ocular lubricants were measured to evaluate the correlation between viscosity and NIKBUT. The complex viscosity of each lubricant was measured three times for each angular frequency (ranging from 0.1 to 100 rad/s) using the Discovery HR-2 hybrid rheometer. The NIKBUT measurements were performed eight times for each lubricant using an advanced eye model mounted on the OCULUS Keratograph 5M. A contact lens (CL; ACUVUE OASYS [etafilcon A]) or a collagen shield (CS) was used as the simulated corneal surface. Phosphate-buffered saline was used as a simulated fluid.
Results:
The results showed a positive correlation between viscosity and NIKBUT at high shear rates (at 10 rad/s, r = 0.67) but not at low shear. This correlation was even better for viscosities between 0 and 100 mPa*s (r = 0.85). Most of the lubricants tested in this study also had shear-thinning properties. OPTASE INTENSE, I-DROP PUR GEL, I DROP MGD, OASIS TEARS PLUS, and I-DROP PUR had higher viscosity in comparison to other lubricants (P < 0.05). All of the formulations had a higher NIKBUT than the control (2.7 ± 1.2 seconds for CS and 5.4 ± 0.9 seconds for CL) without any lubricant (P < 0.05). I-DROP PUR GEL, OASIS TEARS PLUS, I-DROP MGD, REFRESH OPTIVE ADVANCED, and OPTASE INTENSE had the highest NIKBUT using this eye model.
Conclusions:
The results show that the viscosity is correlated with NIKBUT, but further work is necessary to determine the underlying mechanisms.
Translational Relevance:
The viscosity of ocular lubricants can affect NIKBUT and tear film stability, so it is an important property to consider when formulating ocular lubricants.
More Related Videos
06:29Effect of Artificial Tear Formulations on the Metabolic Activity of Human Corneal Epithelial Cells after Exposure to Desiccation
Published on: May 2, 2020
08:04A Rabbit Model of Aqueous-Deficient Dry Eye Disease Induced by Concanavalin A Injection into the Lacrimal Glands: Application to Drug Efficacy Studies
Published on: January 24, 2020
Related Concept Videos
Surface Tension, Capillary Action, and Viscosity
The various IMFs between identical molecules of a substance are examples of cohesive forces. The molecules within a liquid are surrounded by other molecules and are attracted equally in all directions by the cohesive forces within the liquid. However, the molecules on the surface of a liquid are attracted only by about one-half as many molecules. Because of the unbalanced molecular attractions on the surface molecules, liquids contract to form a shape that minimizes the number...
Viscosity
The SI unit of viscosity is...