An Unusual Cause of End-Tidal Carbon Dioxide Rise During One-Lung Ventilation
Jason T Boyle1, Andre F Gosling1, Benjamin Wei2
1Cardiothoracic Anesthesiology, University of Alabama at Birmingham School of Medicine, Birmingham, USA.
Abstract:
A relatively common problem that may arise during one-lung ventilation is elevation of end-tidal carbon dioxide (ETCO2), which has several potential etiologies. This case report describes a 69-year-old woman with carcinoid tumor undergoing a robotic left lower lobectomy complicated by an acute rise in ETCO2 during one-lung ventilation, without an immediately identifiable cause. Thorough evaluation revealed CO2 leak through an open bronchial lumen resulting in an artificially high ETCO2 reading. This case report demonstrates the importance of performing a comprehensive assessment during acute changes in ETCO2 while also considering changes in the surgical field, which may contribute to these findings.
More Related Videos
Related Concept Videos
Acute Respiratory Failure-III
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes...
Hyperpnea and Hyperventilation
Acute Respiratory Failure-I
Definition: It is defined by specific criteria based on blood gas measurements. Hypoxemia happens when the partial pressure of oxygen (PaO2) falls below 60 mmHg. At the same time,...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include: