Related Experiment Video
Updated: Jul 24, 2025

Determination of Inorganic Arsenic in a Wide Range of Food Matrices using Hydride Generation - Atomic Absorption Spectrometry.
Published on: September 1, 2017
Development of a New Sensitive Method for Lead Determination by Platinum-Coated Tungsten-Coil Hydride Generation
1Muğla Vocational School, Chemistry and Chemical Treatment Technologies Department, Chemistry Technology Program, Muğla Sıtkı Koçman University, 48000 Muğla, Turkey.
Abstract:
A novel very sensitive and rapid analytical method was improved where gaseous lead formed was transported to and trapped on an externally heated platinum-coated tungsten-coil atom trap for in situ preconcentration. The analytical performance of the developed method with the graphite furnace atomic absorption spectrometry (GFAAS) method was compared. All critical parameters affecting the performance of both methods were optimized. The limit of quantitation (LOQ) was found as 11.0 ng L-1 and the precision was 2.3% in terms of percent relative standard deviation (RSD%). Characteristic concentration (Co) of the developed trap method was indicating a 32.5-fold enhancement in sensitivity compared to the GFAAS method. In order to investigate the surface morphology of the W-coil, scanning electron microscope-energy-dispersive X-ray (SEM-EDS) analyzes were performed. The accuracy of the trap method was tested by certified reference materials: NIST SRM 1640a (the elements in natural water) and DOLT:5 (dogfish liver). Interferences from other hydride-forming elements were investigated. Application of the trap method was demonstrated by the analysis of some drinking water and fish tissue samples. The t test was applied to drinking water samples, and the results indicated that there was no statistically significant error.
More Related Videos
10:31Detection and Recovery of Palladium, Gold and Cobalt Metals from the Urban Mine Using Novel Sensors/Adsorbents Designated with Nanoscale Wagon-wheel-shaped Pores
Published on: December 6, 2015
05:35Quantification of Metal Leaching in Immobilized Metal Affinity Chromatography
Published on: January 17, 2020
Related Concept Videos
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Atomic Emission Spectroscopy: Lab
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Absorption Spectroscopy: Atomization Methods
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....