Contamination of the central medical air supply with water leading to mass ventilator failure
J J X Chia1, M H Goh1, M M Goh2
1Department of Anaesthesiology Singapore General Hospital Singapore.
Abstract:
Here, we present a case of mass ventilator failure due to contaminated medical air. Multiple ventilators failed routine tests, including almost all of the ventilators in our intensive care unit. A faulty air compressor had led to water contamination of our centre's supply of medical air. Water entered the pipeline supply of air and, hence the ventilators and anaesthetic machines. The disruption of the machines' proportional mixer valve resulted in unreliable delivery of fresh gas flow. This malfunction was discovered during routine pre-use checks, and backup ventilators were available to replace the faulty ventilators. A shortage of equipment was averted due to a serendipitous availability of ventilator stockpiles prepared for the COVID-19 pandemic. Ventilator shortages are commonly described in mass casualty and pandemic scenarios. While there are multiple strategies described in literature to augment and maximise equipment available for mechanical ventilation, stockpiling equipment remains an expensive but necessary component of disaster contingency planning.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...
Acute Respiratory Failure-III
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Acute Respiratory Failure-I
Definition: It is defined by specific criteria based on blood gas measurements. Hypoxemia happens when the partial pressure of oxygen (PaO2) falls below 60 mmHg. At the same time,...