Spectral metal artifact reduction after posterior spinal fixation in photon-counting detector CT datasets
Daniel Popp1, Andrea X Sinzinger1, Josua A Decker1
1Department of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Augsburg, Stenglinstr. 2, 86156 Augsburg, Germany.
Purpose:
To investigate the usefulness of virtual monoenergetic image (VMI) reconstructions derived from scans on a novel photon-counting detector CT (PCD-CT) for artifact reduction in patients after posterior spinal fixation.
Methods:
This retrospective cohort study included 23 patients status post posterior spinal fixation. Subjects were scanned on a novel PCD-CT (NAEOTOM Alpha, Siemens Healthineers, Erlangen, Germany) as part of routine clinical care. 14 sets of VMI reconstructions were derived in 10 keV increments for the interval 60-190 keV. The mean and the standard deviation (SD) of CT-values in 12 defined locations around a pair of pedicle screws on one vertebral level and the SD of homogenous fat were measured and used to calculate an artifact index (AIx).
Results:
Averaged over all regions, the lowest AIx was observed at VMI levels of 110 keV (32.5 (27.8-37.9)) which was significantly different from those of VMIs ≤ 90 keV (p < 0.001) or ≥160 keV (p < 0.015), respectively. Overall AIx values increased in both lower- and higher-keV levels. Regarding individual locations, either a monotonous AIx-decrease for increasing keV values or an AIx-minimum in intermediate-keV levels (100-140 keV) was found. In locations adjacent to larger metal parts, the increase of AIx values at the high-end of the keV spectrum was mainly explained by a reappearance of streak artifacts.
Conclusion:
Our findings suggest that 110 keV is the optimal VMI setting for overall artifact suppression. In specific anatomical regions, however, slight adjustments towards higher-keV levels may provide better results.