Effects of apparatus dead space on volumetric capnograms in neonates with healthy lungs: a simulation study
Marcelo D Campos1, Lucio H Palazzi2, Stephan H Böhm3
1Department of Anesthesiology, Sanatorio Finochietto, Buenos Aires, Argentina.
Background:
Volumetric capnography in healthy ventilated neonates showed deformed waveforms, which are supposedly due to technological limitations of flow and carbon dioxide sensors.
Aims:
This bench study analyzed the role of apparatus dead space on the shape of capnograms in simulated neonates with healthy lungs.
Methods:
We simulated mechanical breaths in neonates of 2, 2.5, and 3 kg of body weight using a neonatal volumetric capnography simulator. The simulator was fed by a fixed amount of carbon dioxide of 6 mL/kg/min. Such simulator was ventilated in a volume control mode using fixed ventilatory settings with a tidal volume of 8 mL/kg and respiratory rates of 40, 35, and 30 breaths per minute for the 2, 2.5 and 3 kg neonates, respectively. We tested the above baseline ventilation with and without an additional apparatus dead space of 4 mL.
Results:
Simulations showed that adding the apparatus dead space to baseline ventilation increased the amount of re-inhaled carbon dioxide in all neonates: 0.16 ± 0.01 to 0.32 ± 0.03 mL (2 kg), 0.14 ± 0.02 to 0.39 ± 0.05 mL (2.5 kg), and 0.13 ± 0.01 to 0.36 ± 0.05 mL (3 kg); (p < .001). Apparatus dead space was computed as part of the airway dead space, and therefore, the ratio of airway dead space to tidal volume increased from 0.51 ± 0.04 to 0.68 ± 0.06, from 0.43 ± 0.04 to 0.62 ± 0.01 and from 0.38 ± 0.01 to 0.60 ± 0.02 in the 2, 2.5 and 3 kg simulated neonates, respectively (p < .001). Compared to baseline ventilation, adding apparatus dead space decreased the ratio of the volume of phase III to VT size from 31% to 11% (2 kg), from 40% to 16% (2.5 kg) and from 50% to 18% (3 kg); (p < .001).
Conclusions:
The addition of a small apparatus dead space artificially deformed the volumetric capnograms in simulated neonates with healthy lungs.
More Related Videos
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Respiratory Volumes and Capacities
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Respiratory Volumes and Capacities I
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...