Related Experiment Videos

Thromboembolic complications of the Jarvik-7 total artificial heart: case report

Artificial Organs
|June 1, 1986
PubMed

Insights

The Jarvik-7 total artificial heart (TAH) can cause blood clots, leading to stroke. Thrombus formation occurred in specific TAH components, linked to patient hypercoagulability and heparin resistance.

Area of Science:

  • Biomedical Engineering
  • Cardiovascular Surgery
  • Hematology

Background:

  • The Jarvik-7 total artificial heart (TAH) is a device for end-stage heart failure.
  • Cerebral embolic events are a significant complication in TAH recipients.

Observation:

  • Examination of an explanted Jarvik-7 TAH revealed thrombus formation.
  • Thrombotic deposition was concentrated in crevices within the aortic quick connector and around the valves.

Findings:

  • Thrombus formation was observed in the aortic quick connector and around all four valves.
  • The most severe deposition was on the atrioventricular valves' outflow surface and around the pulmonic valve crevices.
  • The patient exhibited developing hypercoagulability and increasing resistance to heparin, correlating with the embolic event.

Implications:

  • Understanding thrombus formation sites is crucial for TAH design improvements.
  • Patient-specific hypercoagulability and heparin resistance may increase embolic risk.
  • Further research is needed to mitigate thrombotic complications in TAH patients.