Plan Assessment Metrics for Dose Painting in Stereotactic Radiosurgery
Benjamin Z Tham1, Dionne Aleman1, Håkan Nordström2
1Department of Mechanical and Industrial Engineering, University of Toronto, Toronto, Ontario, Canada.
Advances in Radiation Oncology
|July 7, 2023
Summary
New metrics are proposed to assess the quality of dose painting stereotactic radiosurgery (SRS) plans. These metrics are tailored for nonuniform radiation doses, addressing limitations of existing methods for SRS treatments.
Area of Science:
- Radiation Oncology
- Medical Physics
- Image-Guided Therapy
Background:
- Advancements in radiation therapy precision enable dose painting, delivering nonuniform radiation doses to targets.
- Stereotactic radiosurgery (SRS) is well-suited for dose painting due to its high precision.
- Existing dose painting metrics are not optimized for SRS, which prioritizes avoiding target underdose and considers single prescriptions.
Purpose of the Study:
- To develop and propose a novel set of metrics for assessing dose painting SRS treatment plans.
- To address the clinical need for metrics that accommodate nonuniform dose prescriptions in SRS.
- To create metrics suitable for SRS plans that avoid target underdose and consider dose fall-off.
Main Methods:
- Dose painting SRS prescriptions were generated using Gamma Knife SRS cases and MRI data.
- Treatment plans were optimized using semi-infinite linear programming.
- Existing SRS metrics were modified, and new metrics like integral dose ratio were proposed and evaluated.
Main Results:
- Modified conformity and efficiency indices were found to be suitable replacements for existing SRS metrics in dose painting scenarios.
- A modified conformity index is appropriate for integral or maximum boost methods in dose painting SRS.
- Modified efficiency indices can effectively replace the existing gradient index for dose painting SRS.
Conclusions:
- The proposed modified SRS metrics accurately assess the quality of dose painting SRS plans.
- These new metrics provide equivalent values to original SRS metrics when applied to single-prescription plans.
- The developed metrics meet clinical needs for evaluating complex, nonuniform dose distributions in SRS.