A new model to accurately measure the radon exhalation rate from soil using AlphaGUARD
Shuaibin Liu1, Tao Hu1, Fen Lin1
1College of Physics and Electronic Engineering, Hengyang Normal University, Hengyang, Hunan Province, 421008, China.
Abstract:
Radon exhalation rate from soil is a critical factor in evaluating environmental radon levels. However, AlphaGUARD PQ2000PRO may have some sensitivity towards thoron, which can have a significant impact on radon measurement. The traditional radon exhalation models generally ignore the presence of thoron, leading to an overestimation of the radon exhalation rate from soil. To handle this issue, a new model was proposed based on an analysis of several previous studies on radon exhalation theories. To prove the feasibility of the model, the radon exhalation rate measurements were performed by two different types of detectors-AlphaGUARD PQ2000PRO and RAD7. The radon exhalation rate obtained by using the new model is in good agreement with that obtained by using the theoretical model of radon exhalation of RAD7 within one standard error. This new model can be applied to accurately measure radon exhalation rate from soil by the PIC detector (PQ2000PRO).