A dual-centre study on the radioprotective effect of a novel X-ray protection device during coronary intervention
1Department of Cardiology, First Affiliated Hospital of Nanjing Medical University, Nanjing, Jiangsu, China.
Aim:
To investigate the shielding efficiency of a novel X-ray protection device (NPD) compared with the traditional lead clothing (TLC) during coronary intervention.
Materials And Methods:
This study was performed prospectively in two centres. A total of 200 coronary interventions were included and assigned equally into the NPD or TLC group. The NPD is a floor-standing X-ray protection device, which mainly composes of a barrel-like frame and two layers of lead rubber. Thermoluminescent dosimeters (TLDs) were adopted to detect the cumulative absorbed doses, and were attached outside the NPD or TLC or body of the first operator at four different height levels in four directions during the procedure.
Results:
The cumulative doses outside the NPD were comparable to that of the TLC (2,398.33 ± 2,341.64 versus 1,624.09 ± 1,732.20 μSv, p=0.366), and the cumulative doses inside the NPD were significantly lower than those inside the TLC (40 ± 0 versus 732.28 ± 919.83 μSv, p<0.001). As the TLC did not cover the calf segment of the operator, the area at 50 cm height from the floor in the TLC group was unshielded. The shielding efficiency of NPD was significantly higher than that of the TLC (98.2 ± 0.63% versus 52.11 ± 38.97%, p=0.021).
Conclusion:
The NPD has a significantly higher shielding efficacy than that of the TLC, in particular, it protects the operators' lower limb, liberates their lower body from wearing heavy lead apron, and may consequently reduce the radiation or body-load associated complications.
More Related Videos
06:39Novel Percutaneous Approach for Deployment of 3D Printed Coronary Stenosis Implants in Swine Models of Ischemic Heart Disease
Published on: February 18, 2020
05:07Author Spotlight: Improved Localization and Monitoring of Coronary Flow Reserve Using Modified PLAX View in Mice
Published on: August 25, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Coronary Artery Disease IV: Preventive Measures
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Coronary Artery Disease V: Interprofessional Care
Radiological Investigation I: X-ray and CT