Related Experiment Video
Updated: Jul 24, 2025

A Reversible, Non-invasive Method for Airway Resistance Measurements and Bronchoalveolar Lavage Fluid Sampling in Mice
Published on: April 13, 2010
A novel method for assessment of airway opening pressure without the need for low-flow insufflation
Anne-Fleur Haudebourg1,2, Elsa Moncomble1,2, Arnaud Lesimple3,4
1Assistance Publique-Hôpitaux de Paris, CHU Henri Mondor-Albert Chenevier, Service de Médecine Intensive Réanimation, 51, Avenue du Maréchal de Lattre de Tassigny, 94010, Créteil Cedex, France.
Background:
Airway opening pressure (AOP) detection and measurement are essential for assessing respiratory mechanics and adapting ventilation. We propose a novel approach for AOP assessment during volume assist control ventilation at a usual constant-flow rate of 60 L/min.
Objectives:
To validate the conductive pressure (Pcond) method, which compare the Pcond-defined on the airway pressure waveform as the difference between the airway pressure level at which an abrupt change in slope occurs at the beginning of insufflation and PEEP-to resistive pressure for AOP detection and measurement, and to compare its respiratory and hemodynamic tolerance to the standard low-flow insufflation method.
Methods:
The proof-of-concept of the Pcond method was assessed on mechanical (lung simulator) and physiological (cadavers) bench models. Its diagnostic performance was evaluated in 213 patients, using the standard low-flow insufflation method as a reference. In 45 patients, the respiratory and hemodynamic tolerance of the Pcond method was compared with the standard low-flow method.
Measurements And Main Results:
Bench assessments validated the Pcond method proof-of-concept. Sensitivity and specificity of the Pcond method for AOP detection were 93% and 91%, respectively. AOP obtained by Pcond and standard low-flow methods strongly correlated (r = 0.84, p < 0.001). Changes in SpO2 were significantly lower during Pcond than during standard method (p < 0.001).
Conclusion:
Determination of Pcond during constant-flow assist control ventilation may permit to easily and safely detect and measure AOP.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Suctioning the Nasopharyngeal Airway
Equipment Required
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...