Calibration of Vector Network Analyzer for Measurements in Radio Frequency Propagation Channels
Published on: June 2, 2020
Meniscus-Corrected Method for Broadband Liquid Permittivity Measurements with an Uncalibrated Vector Network Analyzer
Michał Kalisiak1, Arkadiusz Lewandowski1, Wojciech Wiatr1
1Institute of Electronic Systems, The Faculty of Electronics and Information Technology, Warsaw University of Technology, Nowowiejska 15/19, 00-665 Warsaw, Poland.
This study introduces a new calibration-independent method for broadband permittivity characterization of liquids. It accurately measures liquid permittivity by mathematically removing systematic errors, including meniscus effects, reducing calibration costs.
Area of Science:
- Electrical Engineering
- Materials Science
- Physical Chemistry
Background:
- Accurate broadband permittivity characterization of liquids is crucial for various scientific and industrial applications.
- Traditional methods often require complex calibration procedures and specialized equipment, increasing costs and complexity.
- The presence of a meniscus in semi-open test cells introduces significant systematic errors in measurements.
Purpose of the Study:
- To develop a novel, calibration-independent method for broadband permittivity characterization of liquids.
- To address and mathematically remove systematic errors caused by uncalibrated vector network analyzers and liquid meniscus effects.
- To reduce the financial and labor costs associated with system calibration for liquid permittivity measurements.
Main Methods:
- Utilizing three scattering matrix measurements at different liquid levels within a semi-open vertically oriented test cell.
- Applying mathematical operations to isolate and eliminate systematic errors from the vector network analyzer and the liquid meniscus.
- Employing a calibration-independent approach, eliminating the need for expensive standards and skilled labor.
Main Results:
- The novel method demonstrates comparable results to existing calibration-dependent methods for propan-2-ol (IPA) and its aqueous solutions.
- The method successfully removes systematic errors, including those caused by the meniscus, a first for calibration-independent techniques.
- Some challenges were observed when measuring high-loss liquid samples, such as distilled water.
Conclusions:
- The developed method offers a cost-effective and efficient alternative for broadband permittivity characterization of liquids.
- It significantly reduces expenditures related to system calibration, skilled labor, and expensive standards.
- Further refinement may be needed to address limitations with high-loss dielectric materials.