Image Acquisition Method for the Sonographic Assessment of the Inferior Vena Cava
Published on: January 13, 2023
Inferior vena cava distensibility from subcostal and trans-hepatic imaging using both M-mode or artificial
Filippo Sanfilippo1,2, Luigi La Via3,4, Veronica Dezio4
1Department of Anaesthesia and Intensive Care, A.O.U. Policlinico-San Marco, Site "Policlinico G. Rodolico", Via S. Sofia N 78, 95123, Catania, Italy. filipposanfi@yahoo.it.
Background:
Variation of inferior vena cava (IVC) is used to predict fluid-responsiveness, but the IVC visualization with standard sagittal approach (SC, subcostal) cannot be always achieved. In such cases, coronal trans-hepatic (TH) window may offer an alternative, but the interchangeability of IVC measurements in SC and TH is not fully established. Furthermore, artificial intelligence (AI) with automated border detection may be of clinical value but it needs validation.
Methods:
Prospective observational validation study in mechanically ventilated patients with pressure-controlled mode. Primary outcome was the IVC distensibility (IVC-DI) in SC and TH imaging, with measurements taken both in M-Mode or with AI software. We calculated mean bias, limits of agreement (LoA), and intra-class correlation (ICC) coefficient.
Results:
Thirty-three patients were included. Feasibility rate was 87.9% and 81.8% for SC and TH visualization, respectively. Comparing imaging from the same anatomical site acquired with different modalities (M-Mode vs AI), we found the following IVC-DI differences: (1) SC: mean bias - 3.1%, LoA [- 20.1; 13.9], ICC = 0.65; (2) TH: mean bias - 2.0%, LoA [- 19.3; 15.4], ICC = 0.65. When comparing the results obtained from the same modality but from different sites (SC vs TH), IVC-DI differences were: (3) M-Mode: mean bias 1.1%, LoA [- 6.9; 9.1], ICC = 0.54; (4) AI: mean bias 2.0%, LoA [- 25.7; 29.7], ICC = 0.32.
Conclusions:
In patients mechanically ventilated, AI software shows good accuracy (modest overestimation) and moderate correlation as compared to M-mode assessment of IVC-DI, both for SC and TH windows. However, precision seems suboptimal with wide LoA. The comparison of M-Mode or AI between different sites yields similar results but with weaker correlation. Trial registration Reference protocol: 53/2022/PO, approved on 21/03/2022.
Related Concept Videos
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...