Related Experiment Video
Updated: Jul 24, 2025

Multimodal Volumetric Retinal Imaging by Oblique Scanning Laser Ophthalmoscopy oSLO and Optical Coherence Tomography OCT
Published on: August 4, 2018
Impact of A-Scan Rate on Image Quality and Acquisition Time in OCT
Jan-Philipp Bodenbender1, Martin Kowalski1, Katarina Stingl1
1Department for Ophthalmology, University of Tuebingen, Tübingen, Germany.
Purpose:
This study aims to examine the impact of the A-scan rate in optical coherence tomography (OCT) on scan quality and acquisition time.
Methods:
Two horizontal OCT scans per scan rate (20, 85, 125 kHz) of the right eye were captured with the same OCT device (Spectralis SHIFT, HRA + OCT, Heidelberg Engineering GmbH, Heidelberg, Germany) of patients who presented to the inherited retinal dystrophies consultation, thus predominantly challenging patients due to reduced fixation ability. Scan quality was measured by the Q score, a signal-to-noise-ratio (SNR). Acquisition time was measured in seconds.
Results:
Fifty-one patients were included in the study. The highest quality was seen for an A-scan rate of 20 kHz (44.49 dB), succeeded by scans of an A-Scan rate of 85 kHz (38.53 dB) and of 125 kHz (36.65 dB). Differences in scan quality between the various A-scan rates were statistically significant. The acquisition time seen for an A-scan rate of 20 kHz (6.45 s) was significantly longer than those seen for an A-Scan rate of 85 kHz (1.51 s) and of 125 kHz (1.69 s).
Conclusion:
An A-scan rate of 20 kHz resulted in a significantly higher scan quality but also a significantly longer acquisition time compared to scan rates of 85 kHz and 125 kHz. Differences between an A-scan rate of 85 kHz and 125 kHz were marginal.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography