Does cone-beam CT convex triangular field of view influence the image shape distortion of high-density materials?
Deivi Cascante-Sequeira1, Fernanda Coelho-Silva1,2, Lucas P Lopes Rosado1,3
1Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, University of Campinas, Piracicaba, Brazil.
Objectives:
To assess the distortion of high-density materials using two CBCT devices presenting convex triangular and cylindrical fields of view (FOVs).
Methods And Materials:
Four high-density cylinders were individually placed in a polymethylmethacrylate phantom. 192 CBCT scans were acquired using the convex triangular and cylindrical FOVs of Veraviewepocs® R100 (R100) and Veraview® X800 (X800) devices. Using HorosTM's software, two oral radiologists determined the cylinders' horizontal and vertical dimensional alterations. Nine oral radiologists subjectively identified each cylinder's axial shape distortion. Statistical analysis comprised Multiway ANOVA (α = 5%), and the Kruskal-Wallis test.
Results:
The distortion in the axial plane was greater in the convex triangular FOVs for both devices in almost all the materials (p < 0.05). The evaluators subjectively identified a shape distortion in both FOVs for R100 device (p < 0.001), while no distortion was identified for X800 device (p = 0.620). A vertical magnification of all materials was observed in both FOVs for both devices (p < 0.05). No differences among vertical regions (p = 0.988) nor FOVs (p = 0.544) were found for the R100 device, while all materials showed higher magnification in all regions in the cylindrical FOV (p < 0.001) of the X800 device.
Conclusions:
The convex triangular FOV influenced the axial distortion of the high-density materials in both devices. A vertical magnification was observed in both FOVs of both devices, but it was greater in the cylindrical FOV of the X800 device.
Related Concept Videos
Deformations in a Transverse Cross Section
As the material stretches, it expands or contracts in orthogonal directions to the load. This phenomenon varies...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Three-Dimensional Analysis of Strain