Impact of filter configurations on bipolar EGMs: An optimal filter setting for identifying VT substrates
Masateru Takigawa1,2,3,4, Frederic Sacher1,2, Claire Martin1,5,6
1Department of Cardiac Pacing and Electrophysiology, Bordeaux University Hospital (CHU), Bordeaux, France.
Background:
The impact of filtering on bipolar electrograms (EGMs) has not been systematically examined. We tried to clarify the optimal filter configuration for ventricular tachycardia (VT) ablation.
Methods:
Fifteen patients with VT were included. Eight different filter configurations were prospectively created for the distal bipoles of the ablation catheter: 1.0-250, 10-250, 100-250, 30-50, 30-100, 30-250, 30-500, and 30-1000 Hz. Pre-ablation stable EGMs with good contact (contact force > 10 g) were analyzed. Baseline fluctuation, baseline noise, bipolar peak-to-peak voltage, and presence of local abnormal ventricular activity (LAVA) were compared between different filter configurations.
Results:
In total, 2276 EGMs with multiple bipolar configurations in 246 sites in scar and border areas were analyzed. Baseline fluctuation was only observed in the high-pass filter of (HPF) ≤ 10 Hz (p < .001). Noise level was lowest at 30-50 Hz (0.018 [0.012-0.029] mV), increased as the low-pass filter (LPF) extended, and was highest at 30-1000 Hz (0.047 [0.041-0.061] mV) (p < .001). Conversely, the HPF did not affect the noise level at ≤30 Hz. As the HPF extended to 100 Hz, bipolar voltages significantly decreased (p < .001), but were not affected when the LPF was extended to ≥100 Hz. LAVAs were most frequently detected at 30-250 Hz (207/246; 84.2%) and 30-500 Hz (208/246; 84.6%), followed by 30-1000 Hz (205/246; 83.3%), but frequently missed at LPF ≤ 100 Hz or HPF ≤ 10 Hz (p < .001). A 50-Hz notch-filter reduced the bipolar voltage by 43.9% and LAVA-detection by 34.5% (p < .0001).
Conclusion:
Bipolar EGMs are strongly affected by filter settings in scar/border areas. In all, 30-250 or 30-500 Hz may be the best configuration, minimizing the baseline fluctuation, baseline noise, and detecting LAVAs. Not applying the 50-Hz notch filter may be beneficial to avoid missing VT substrate.
More Related Videos
08:22Author Spotlight: Advancing the Study of Brain-Heart Interplay with a Comprehensive EEGLAB Plugin for Multimodal Signal Analysis
Published on: April 26, 2024
09:36Dual-Dye Optical Mapping of Hearts from RyR2R2474S Knock-In Mice of Catecholaminergic Polymorphic Ventricular Tachycardia
Published on: December 22, 2023
Related Concept Videos
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias III: Characteristics of Dysrhythmias
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Dysrhythmias II: Classification of Tachyarrhythmias
Bode Plots Construction
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...