A Detailed Protocol for Physiological Parameters Acquisition and Analysis in Neurosurgical Critical Patients
Published on: October 17, 2017
The Effect of Different Optic Nerve Sheath Diameter Measurements Using Ultrasound to Assess Intracranial Pressure in
Jun Tian1, Guo-Biao Wu, Xiao-Bing Liu
1Department of Neurosurgery (JT), Shijiazhuang People's Hospital, Shijiazhuang, Hebei, China; Department of Neurosurgery (G-BW), The Second Hospital of Hebei Medical University, Shijiazhuang, Hebei, China; and Departments of Neurosurgery (X-BL) and Critical Care Medicine (Z-YW, J-YG), The Third Hospital of Hebei Medical University, Shijiazhuang, Hebei, China.
Background:
Optic nerve sheath diameter (ONSD) is a promising, noninvasive invasive intracranial pressure (ICP) measurement method. This study aims to analyze the differences in ONSD between the left and right eyeballs and the differences in ultrasonic measurement between the transverse and sagittal planes.
Methods:
Data from a total of 50 eligible patients with various types of brain injury who were admitted to our hospital from May 2019 to June 2021 were analyzed. An ONSD assessment was then performed using Philips B-mode ultrasound, measuring ONSD 3 mm posterior to the eyeballs. The left and right ONSDs in the transverse and sagittal planes were measured. Intraparenchymal fiber optic sensors and catheters were inserted into the ventricles and connected to an external pressure transducer to measure ICP.
Results:
A total of 164 sonographic measurements of ONSD were performed in 50 patients with brain injury in a prospective observational study. Statistically significant differences were found in ONSD between the transverse and sagittal planes. The difference in the left ONSD between the transverse and sagittal planes was 0.007 ± 0.030 cm ( P = 0.003). The Spearman rank correlation test showed that the correlation coefficient between ICP and left/right ONSD in the transverse/sagittal planes was 0.495 vs 0.546 and 0.559 vs 0.605, respectively. The results showed that the areas under the curve of ONSD in the transverse and sagittal planes were 0.843 and 0.805, respectively. Medcalc software was used to compare the areas under the receiver operator characteristic curve, and the results showed that ONSD in the sagittal plane is generally better than in the transverse plane ( P = 0.0145).
Conclusions:
This study found that ONSD in the sagittal plane is superior to the transverse plane regarding the comprehensive efficacy of ICP, and unilateral measurement is sufficient.