Impact of Dual-Energy Computed Tomography (DECT) Postprocessing Protocols on Detection of Monosodium Urate (MSU)
Andrea Sabine Klauser1, Sylvia Strobl1, Christoph Schwabl1
1Department for Radiology, Medical University of Innsbruck, 6020 Innsbruck, Austria.
Objective:
To evaluate two different dual-energy computed tomography (DECT) post-processing protocols for the detection of MSU deposits in foot tendons of cadavers with verification by polarizing light microscopy as the gold standard.
Material And Methods:
A total of 40 embalmed cadavers (15 male; 25 female; median age, 82 years; mean, 80 years; range, 52-99; SD ± 10.9) underwent DECT to assess MSU deposits in foot tendons. Two postprocessing DECT protocols with different Hounsfield unit (HU) thresholds, 150/500 (=established) versus 120/500 (=modified). HU were applied to dual source acquisition with 80 kV for tube A and 140 kV for tube B. Six fresh cadavers (4 male; 2 female; median age, 78; mean, 78.5; range 61-95) were examined by DECT. Tendon dissection of 2/6 fresh cadavers with positive DECT 120 and negative DECT 150 studies were used to verify MSU deposits by polarizing light microscopy.
Results:
The tibialis anterior tendon was found positive in 57.5%/100% (DECT 150/120), the peroneus tendon in 35%/100%, the achilles tendon in 25%/90%, the flexor halluces longus tendon in 10%/100%, and the tibialis posterior tendon in 12.5%/97.5%. DECT 120 resulted in increased tendon MSU deposit detection, when DECT 150 was negative, with an overall agreement between DECT 150 and DECT 120 of 80% (p = 0.013). Polarizing light microscope confirmed MSU deposits detected only by DECT 120 in the tibialis anterior, the achilles, the flexor halluces longus, and the peroneal tendons.
Conclusion:
The DECT 120 protocol showed a higher sensitivity when compared to DECT 150.
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...