Related Experiment Video
Updated: Jul 23, 2025

Oral Biofilm Formation on Different Materials for Dental Implants
Published on: June 24, 2018
Investigation of the Influence of Roughness and Dental Implant Design on Primary Stability via Analysis of Insertion
Marta Romero1, Mariano Herrero-Climent1,2, Blanca Ríos-Carrasco1
1Department of Periodontology, School of Dentistry, Universidad de Seville, C/Avicena S/N, 41009 Seville, Spain.
Abstract:
In the placement of dental implants, the primary fixation between the dental implant and the bone is of great importance and corresponds to compressive mechanical fixation that aims to prevent micromovement of the implant. The aim of this research was to determine the role of roughness and the type of dental implant (tissue-level or bone-level) in implant stability, measured using resonance frequency analysis (RFA) and insertion torque (IT). We analyzed 234 titanium dental implants, placed in fresh calf ribs, at the half-tissue level and half-bone level. The implant surface was subjected to grit-blasting treatments with alumina particles of 120, 300, and 600 μm at a projection pressure of 2.5 bar, resulting in three types of roughness. Roughness was determined via optical interferometry. The wettability of the surfaces was also determined. Implant stability was measured using a high-precision torquemeter to obtain IT, and RFA was used to determine the implant stability quotient (ISQ). The results show that rough surfaces with Sa values of 0.5 to 4 μm do not affect the primary stability. However, the type of implant is important; bone-level implants obtained the highest primary stability values. A good correlation between the primary stability values obtained via IT and ISQ was demonstrated. New in vivo studies are necessary to know whether these results can be maintained in the long term.

