Related Experiment Video
Updated: Jul 23, 2025

Determination of Inorganic Arsenic in a Wide Range of Food Matrices using Hydride Generation - Atomic Absorption Spectrometry.
Published on: September 1, 2017
Determination of Arsenic Species in Edible Insects by LC-ICP-MS
Eri Matsumoto1, Makoto Matsumoto1
1Japan Food Research Laboratories, 6-11-10 Nagayama, Tama-shi, Tokyo 206-0025, Japan.
Background:
Edible insects may contain arsenic. Analysis of arsenic species is necessary in order to accurately assess arsenic exposure.
Objective:
An analytical method was validated and used to determine and quantitate arsenic species in edible insects.
Methods:
Arsenic species were extracted from edible insects by heating at 100°C in 0.3 mol/L nitric acid. The concentration of arsenic species was then determined by LC-inductively coupled plasma-mass spectrometry (LC-ICP-MS) using an octadecylsilane (ODS) column with a mobile phase containing an ion-pair reagent.
Results:
The LOD (0.007-0.012 mg/kg), LOQ (0.021-0.038 mg/kg), repeatability (1.2-3.2%), intermediate precision (2.8-4.5%), and trueness (recoveries 97-102% based on spiked samples) of the proposed method were satisfactory for inorganic arsenic, dimethylarsinic acid (DMA), and arsenobetaine (AB) in edible insects. Total arsenic was detected in all samples obtained in Japan (Asian forest scorpion, diving beetles, giant water bug, grasshoppers, June beetles, mole crickets, male rhino beetle, female rhino beetle, sago worms, and silkworm pupae) and consisted of mostly inorganic arsenic. Beetles in particular showed relatively high levels.
Conclusion:
Arsenic content varies among edible insect species. Feed control is important, as arsenic concentrations in edible insects may be feed dependent.
Highlights:
Arsenic species in edible insects were analyzed by LC-ICP-MS using an ODS column with a mobile phase containing an ion-pair reagent. Inorganic arsenic was detected in most samples, with concentrations ranging from <0.04 to 29.3 mg/kg.
More Related Videos
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
08:01Removal of Arsenic Using a Cationic Polymer Gel Impregnated with Iron Hydroxide
Published on: June 28, 2019
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...