Coaxial Placement of Balloon-Expandable and Self-Expanding Stents: Impact on Crush Resistance and Luminal Recovery in
Ningcheng Li1, Ramsey Al-Hakim2, Steven Lewis3
1The Charles T. Dotter Department of Interventional Radiology, Oregon Health & Science University, Portland, Oregon.
Abstract:
During endovascular interventions, coaxial deployment of stents may be required to preserve luminal gain. This study characterized in vitro the effect on crush resistance and postcompression recovery when 316L stainless steel balloon-expandable (BE) and laser-cut nitinol self-expanding (SE) venous stents were deployed coaxially. Various stent configurations were parallel-plate compressed from a fully expanded state to 50% diameter reduction (Criterion, Model 42; MTS, Eden Prairie, Minnesota) in a 37 °C ± 1 water bath. Coaxial deployments of SE stent inside BE stent and BE stent inside SE stent demonstrated higher crush resistances compared with each stent individually or their mathematical summation (analysis of variance P < .0001; pairwise comparison P < .01). The configuration of SE stent inside BE stent showed higher postcompression luminal recovery at 48.7% compared with that of BE stent inside SE stent at 27.5% (P = .0001). Coaxial deployment of SE stent inside BE stent may improve crush resistance and luminal recovery after compression in the appropriate clinical context.