Related Experiment Video
Updated: Jul 23, 2025

06:54
Author Spotlight: Advancements in Impedance Monitoring for Cochlear Implant Surgery
Published on: August 4, 2023
1.2K
The experience of device failure after cochlear implantation
Jeong Heon Kim1, Yeonjoo Choi1, Woo Seok Kang1
1Department of Otorhinolaryngology-Head and Neck Surgery, Asan Medical Center, University of Ulsan College of Medicine, Seoul, South Korea.
Summary
Intermittent loss of signal is a key symptom of cochlear implant (CI) device failure. Early electrical examination for this symptom can improve audiological outcomes after revision surgery.
Area of Science:
- Otolaryngology
- Biomedical Engineering
- Audiology
Background:
- Cochlear Implantation (CI) is a vital treatment for severe to profound hearing loss.
- Device failure can significantly impact patient outcomes and quality of life.
- Identifying early signs of CI device failure is crucial for timely intervention.
Purpose of the Study:
- To describe the management of cochlear implant device failures at a tertiary center.
- To identify prodromal symptoms indicative of cochlear implant device failure.
- To analyze audiological outcomes following revision surgeries for device failure.
Main Methods:
- Retrospective review of 1431 cochlear implant surgeries performed between January 2000 and May 2017.
- Analysis of factors including etiology of hearing loss, age at surgery, device type, and symptoms of device failure.
- Evaluation of audiological outcomes using Categories of Auditory Performance (CAP) scores.
Main Results:
- A total of 27 (1.9%) patients experienced device failure requiring revision surgery.
- The most common etiologies were idiopathic hearing loss and cochlear hypoplasia.
- Intermittent loss of signal was the most suggestive symptom of device failure. Advanced Bionics devices showed a higher failure rate (10.0%) compared to Cochlear (1.3%) and MED-EL (3.1%).
- Mean CAP scores improved post-revision surgery, indicating continuous development of auditory performance.
Conclusions:
- Intermittent loss of signal strongly suggests cochlear implant device failure and warrants electrical examination.
- Revision surgeries for device failure can lead to sustained improvement in audiological outcomes.
- Understanding device failure rates across manufacturers is important for patient counseling and device selection.