Background:
Implant-based breast reconstruction (IBBR) can be performed using a variety of biological and synthetic meshes. However, there has yet to be a consensus on the optimal mesh. This study investigates the safety and patient satisfaction of using TiLOOP® Bra in IBBR and compares its postoperative complication risk with that of porcine acellular dermal matrix (ADM) and SERAGYN® BR.
Methods:
The literature review was performed via PRISMA criteria, 23 studies met the inclusion criteria for the TiLOOP® Bra review, and 5 studies met the inclusion criteria for the meta-analysis. Patient characteristics and per-breast complications were collected. Data were analyzed using Cochrane RevMan and IBM SPSS.
Results:
In 3175 breasts of 2685 patients that underwent IBBR using TiLOOP® Bra, rippling was observed as the most common complication, followed by seroma and capsular contracture. No significant difference in the overall complication rate between pre- and sub-pectoral IBBR using TiLOOP® Bra. However, the meta-analysis showed that the TiLOOP® Bra group had significantly lower odds of implant loss, seroma, wound dehiscence, and the need for reoperation or hospitalization than the ADM group. Additionally, the TiLOOP® Bra group had a significantly lower seroma rate compared to the SERAGYN® BR group, while the other outcome indicators were similar between the two groups.
Conclusion:
TiLOOP® Bra has become increasingly popular in IBBR in recent years. This review and meta-analysis support the favorable safety profile of TiLOOP® Bra reported in the current literature. The meta-analysis revealed that TiLOOP® Bra has better safety than ADM and a comparable risk of complications compared to SERAGYN® BR. However, as most studies had low levels of evidence, further investigations are necessary.
Level Of Evidence Iii: